Electron radiotherapy: a study on dosimetric uncertainty using small cutouts

  • 1. Department of Radiation Oncology, University of Toronto and Radiation Medicine Program, Princess Margaret Hospital, University Health Network, 610 University Avenue, Toronto, ON M5G 2M9 (Canada)
  • 2. Medical Physics Department, Grand River Regional Cancer Center, Grand River Hospital, 835 King Street West, Kitchener, ON N2G 1G3 (Canada)

Description

This note investigated the dosimetric uncertainties due to the positional error when centring a small cutout to the machine central beam axis (CAX) in electron radiotherapy. A group of six circular cutouts with 4 cm diameter were made with their centres shifting 0, 2, 4, 6, 8 and 10 mm from the machine CAX for the 6 x 6 cm2 applicator. The per cent depth doses, beam profiles and output factors were measured using the 4, 9 and 16 MeV clinical electron beams produced by a Varian 21 EX linear accelerator. The 2D isodose distributions in the z-x (or cross-line) and z-y (or in-line) plane were calculated by Monte Carlo simulation using the EGSnrc system. When the cutout centre was shifted away from the machine CAX for the 4 MeV beam, the dm, R80 and R90 at the machine CAX had no significant change (<0.1 mm). For higher energies of 9 and 16 MeV beams, the dm was reduced by 0.45 and 1.63 mm per mm, between the cutout centre and the machine CAX with off-axis shift <6 mm respectively. R80 and R90 were reduced by more than 0.3 mm per mm off-axis shift for both energies. The isodose coverage of the in-line axis beam profile was reduced when the cutout centre was shifted away from machine CAX. It is important for oncology staff to note such dosimetric changes in the clinical electron radiotherapy, particularly when a high energy electron beam is used for small cutout. Such positional uncertainty is unavoidable in fabricating an electron cutout in the mould room. (note)

Additional details

Identifiers

DOI
10.1088/0031-9155/52/1/N01;
PII
S0031-9155(07)28295-8;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
52
Journal Issue
1
Journal Page Range
p. N1-N11
ISSN
0031-9155
CODEN
PHMBA7