Published July 1989 | Version v1
Report Open

A neutron detector using silicon pin photodiodes

  • 1. Nagoya Univ. (Japan). Faculty of Engineering

Description

Silicon pin photodiodes in recent years have been used as detectors for charged particle and low energy X- and gamma-ray spectroscopy instead of conventional surface barrier detectors. This is because pin photodiodes have a higher energy resolution, and the cost required for these purposes is lower. The present paper describes the development of a highly sensitive and small sized neutron detector for personal monitoring using pin photodiodes with a gadolinium-foil neutron converter. A detection efficiency of 3.1 % is obtained for thermal neutrons by detecting internal conversion electrons emitted from a gadolinium-foil converter. Gamma-component contained in the output pulses is cancelled nearly completely by using twin photodiode with tin foil instead of the gadolinium converter. A low detection limit of 240 cm-2sec-1 for thermal neutron flux density is obtained under gamma-background with a dose rate of 25 μSv/hr. Since the present neutron detector is also sensitive to gamma-rays, it is suitable to measure the equivalent dose rate in a mixed field of neutrons and gamma-rays. The detector could be useful for the construction of real-time neutron dosimeters. (N.K.)

Files

21047330.pdf

Files (146.4 kB)

Name Size Download all
md5:e438075489cb19251bf108d661627dba
146.4 kB Preview Download
Part of:
Radiation detectors and their uses

Additional details

Publishing Information

Imprint Title
Radiation detectors and their uses
Imprint Pagination
151 p.
Journal Page Range
p. 126-133.
Report number
KEK--89-5

Conference

Title
4. workshop on radiation detectors and their uses.
Dates
24-25 Jan 1989.
Place
Tsukuba, Ibaraki (Japan).

Optional Information