Trouble in source driving system of a 60Co teletherapy unit
Description
An asymmetry in dose profile of a 60Co teletherapy unit was found by means of water-phantom measurement. The reason of that trouble was confirmed to be the abnormal ON position of the source, which is resulted from the high friction between contiguous surfaces of the spring for driving the source to OFF position. Lubrication in the spring improved the mobility a little, but was not a radical repair. The radical repair was to replace the old spring by new one. Periodic maintenance for source driving system and periodic measurement of field aymmetry are required for prevention of abnormal ON position of 60Co source. (Author)