Comparative analysis of partial imaging performance parameters of home and imported X-ray machines
Creators
- 1. Taishan Medical College, Tai'an (China). Dept. of Radiology
Description
Objective: To compare and analyze the performance indexes and the imaging quality of the home and imported X-ray machines through testing their partial imaging performance parameters. Methods: By separate sampling from 10 home and 10 imported X-ray machines, the parameters including tube current, time of exposure, machine total exposure, and repeatability were tested, and the imaging performance was evaluated according to the national standard. Results: All the performance indexes met the standard of GB4505-84. The first sampling tests showed the maximum changing coefficient of imaging performance repeatability of the home X-ray machines was Δmax1 = 0.025, while that of the imported X-ray machine was Δmax1 = 0.016. In the second sampling tests, the maximum changing coefficients of the two were Δmax2 = 0.048 and Δmax2 = 0.022, respectively. Conclusion: The 2 years' follow-up tests indicate that there is no significant difference between the above-mentioned parameters of the elaborately adjusted home X-ray machines and imported ones, but the home X-ray machines are no better than the imported X-ray machines in stability and consistency
Additional details
Publishing Information
- Journal Title
- Chinese Journal of Radiology
- Journal Volume
- 36
- Journal Issue
- 2
- Journal Page Range
- p. 170-173
- ISSN
- 1005-1201
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 34019678
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; PERFORMANCE; QUALITY CONTROL; X-RAY EQUIPMENT; X-RAY RADIOGRAPHY
- Descriptors DEC
- CONTROL; EQUIPMENT; EVALUATION; INDUSTRIAL RADIOGRAPHY; MATERIALS TESTING; NONDESTRUCTIVE TESTING; TESTING