Comparative analysis of partial imaging performance parameters of home and imported X-ray machines

  • 1. Taishan Medical College, Tai'an (China). Dept. of Radiology

Description

Objective: To compare and analyze the performance indexes and the imaging quality of the home and imported X-ray machines through testing their partial imaging performance parameters. Methods: By separate sampling from 10 home and 10 imported X-ray machines, the parameters including tube current, time of exposure, machine total exposure, and repeatability were tested, and the imaging performance was evaluated according to the national standard. Results: All the performance indexes met the standard of GB4505-84. The first sampling tests showed the maximum changing coefficient of imaging performance repeatability of the home X-ray machines was Δmax1 = 0.025, while that of the imported X-ray machine was Δmax1 = 0.016. In the second sampling tests, the maximum changing coefficients of the two were Δmax2 = 0.048 and Δmax2 = 0.022, respectively. Conclusion: The 2 years' follow-up tests indicate that there is no significant difference between the above-mentioned parameters of the elaborately adjusted home X-ray machines and imported ones, but the home X-ray machines are no better than the imported X-ray machines in stability and consistency