Calibration of environmental γ-radiation dosimeters
Description
The reference radiation and procedure for calibration of environmental dosimeters in this laboratory are presented. The total uncertainty of conventional dose value given by MCD-100 type ionization chamber, with air-attenuation and backscattering corrected, is unlikely more than ±2.5% for filtered X-ray, ±1.5% for 137Cs and 60Co γ-radiation respectively
Additional details
Publishing Information
- Journal Title
- Journal of Isotopes
- Journal Volume
- 4
- Journal Issue
- 4
- Journal Page Range
- p. 87-94.
- ISSN
- 1000-7512
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 24007578
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKSCATTERING; CALIBRATION; CALIBRATION STANDARDS; CESIUM 137; COBALT 60; CORRECTIONS; DOSEMETERS; GAMMA DOSIMETRY; IONIZATION CHAMBERS
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; COBALT ISOTOPES; DOSIMETRY; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-EVEN NUCLEI; ODD-ODD NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; SCATTERING; STANDARDS; YEARS LIVING RADIOISOTOPES