Published 2014 | Version v1
Journal article

Development of a measurement device, using a Geiger-Mueller type detector, for the determination of the activity in a 99mTc generator prototype

  • 1. Facultad de Ingenieria de Sistemas y Electronica, Universidad Tecnologica del Peru, Lima (Peru)
  • 2. Direccion de Investigacion y Desarrollo, Instituto Peruano de Energia Nuclear, Lima (Peru)
  • 3. Facultad de Ciencias, Universidad Nacional de Ingenieria, Lima (Peru)
  • 4. Facultad de Quimica e Ingenieria Quimica, Universidad Nacional Mayor de San Marcos, Lima (Peru)

Description

This article presents the implementation of a measurement system using a Geiger-Mueller detector (GM) in order to adapt it into a 99mTc generator prototype. The response signal of the measurement system designed in terms of count rate is linearly proportional to the variation of the activity measured between 280 and 170 mCi of 99mTc with a relative error of ± 2,8 %. However, further tests are needed to evaluate the correlation for an activity level lower than 20 mCi in order to obtain an adequate range of use. (authors).

Additional details

Additional titles

Original title (Spanish)
Desarrollo de un dispositivo de medicion usando un Geiger-Mueller para la determinacion de actividad en un prototipo de generador de 99mTc

Publishing Information

Journal Title
Informe Cientifico Tecnologico (Online)
Journal Volume
14
Series
7 refs., 1 tab, 10 figs.
Journal Page Range
p. 133-136
ISSN