Dilution probe for an aerosol mass monitor

Description

A special probe dilutes high-concentration aerosols to concentrations within the range of the mass monitor, to which it is attached. Tests have established that probe sample losses are less than 2.5% and that sample rates of the system are constant. A preliminary calibration of the system with spent oil shale aerosols yielded a sensitivity of 1 mV per 6 ng of sample collected, whereas the theoretical calibration of the mass monitor is 1 mV per 1.4 ng of sample

Files

10474139.pdf

Files (71.0 kB)

Name Size Download all
md5:f739839f4c094ef9dac2cb8cd6fc9033
71.0 kB Preview Download

Additional details

Publishing Information

Imprint Title
Pacific Northwest Laboratory annual report for 1978 to the DOE Assistant secretary for Environment. Part 1. Biomedical sciences
Journal Page Range
p. 3.5-3.7.
Report number
PNL--2850(Pt.1)

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
10474139
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
AEROSOL MONITORING; AEROSOLS; CALIBRATION; DILUTION; OIL SHALES; PROBES
Descriptors DEC
COLLOIDS; DISPERSIONS; METAMORPHIC ROCKS; MONITORING; ROCKS; SHALES; SOLS