A mathematical expression for %DD accurate from Co-60 to 24 MV
Creators
- 1. University of Minnesota, Department of Therapeutic Radiology-Radiation Oncology, Minneapolis, Minnesota 55455 (USA)
Description
Obtaining accurate %DD values for routine treatment calculations is essential in radiation therapy. Many papers have presented expressions to calculate this or related parameters but these expressions often required that the parameters needed by the equation be determined for each individual treatment unit. This paper presents an expression that calculates %DD values with a mean-square accuracy of ∼1.0% versus measured values. The expression is applicable to beam energies ranging from Co-60 to 24 MV, field sizes from 4x4 to 40x40 cm2, and depths from 1 cm deeper than dmax to 30 cm. The only information required by this expression that is machine specific is the ionization ratio
Additional details
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 18
- Journal Issue
- 4
- Series
- Med. Phys.
- Journal Page Range
- 724-726
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 22084261
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ACCURACY; DEPTH DOSE DISTRIBUTIONS; DOSIMETRY; RADIOTHERAPY
- Descriptors DEC
- DISTRIBUTION; MEDICINE; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS; THERAPY