Published July 1991 | Version v1
Journal article

A mathematical expression for %DD accurate from Co-60 to 24 MV

Creators

  • 1. University of Minnesota, Department of Therapeutic Radiology-Radiation Oncology, Minneapolis, Minnesota 55455 (USA)

Description

Obtaining accurate %DD values for routine treatment calculations is essential in radiation therapy. Many papers have presented expressions to calculate this or related parameters but these expressions often required that the parameters needed by the equation be determined for each individual treatment unit. This paper presents an expression that calculates %DD values with a mean-square accuracy of ∼1.0% versus measured values. The expression is applicable to beam energies ranging from Co-60 to 24 MV, field sizes from 4x4 to 40x40 cm2, and depths from 1 cm deeper than dmax to 30 cm. The only information required by this expression that is machine specific is the ionization ratio

Additional details

Publishing Information

Journal Title
Medical Physics
Journal Volume
18
Journal Issue
4
Series
Med. Phys.
Journal Page Range
724-726
ISSN
0094-2405
CODEN
MPHYA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
22084261
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
ACCURACY; DEPTH DOSE DISTRIBUTIONS; DOSIMETRY; RADIOTHERAPY
Descriptors DEC
DISTRIBUTION; MEDICINE; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS; THERAPY