Published December 1984
| Version v1
Journal article
Exact treatment of the dual-energy method in CT using polyenergetic X-ray spectra
Creators
- 1. Tuebingen Univ. (Germany, F.R.). Medizinisches Strahleninstitut und Roentgenabteilung
Description
The theoretical background of the dual-energy method in CT is discussed in detail, especially the meaning of a measured linear attenuation coefficient of a polyenergetic x-ray spectrum in the presence of beam hardening. Using exact values for the incident spectra the iteration procedure shown here allows one to obtain the true values of the effective atomic number and electron density of an absorber material. The applicability of the method in CT is illustrated by computer simulation of the image reconstruction for a homogeneous and an inhomogeneous phantom. (author)
Additional details
Publishing Information
- Journal Title
- Phys. Med. Biol.
- Journal Volume
- 29
- Journal Issue
- 12
- Series
- Phys. Med. Biol.
- Journal Page Range
- 1501-1510
- ISSN
- 0031-9155
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 16024061
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ATOMIC NUMBER; COMPUTERIZED SIMULATION; COMPUTERIZED TOMOGRAPHY; ELECTRON DENSITY; ENERGY DEPENDENCE; IMAGE PROCESSING; PHANTOMS; X-RAY SPECTRA
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; MOCKUP; SIMULATION; SPECTRA; STRUCTURAL MODELS; TOMOGRAPHY