Development of a positron probe for localization and excision of brain tumours during surgery
Creators
- 1. Laboratoire Imagerie et Modelisation en Neurobiologie et Cancerologie (UMR 8165), Campus d'Orsay, 91406 Orsay Cedex (France)
- 2. Service de neurochirurgie, CHU Henri Mondor, 94010 Creteil Cedex (France) and URA CEA-CNRS 2210, 4 Place du General Leclerc, 91401 Orsay Cedex (France)
Description
The survival outcome of patients suffering from gliomas is directly linked to the complete surgical resection of the tumour. To help the surgeons to delineate precisely the boundaries of the tumour, we developed an intraoperative positron probe with background noise rejection capability. The probe was designed to be directly coupled to the excision tool such that detection and removal of the radiolabelled tumours could be simultaneous. The device consists of two exchangeable detection heads composed of clear and plastic scintillating fibres. Each head is coupled to an optic fibre bundle that exports the scintillating light to a photodetection and processing electronic module placed outside the operative wound. The background rejection method is based on a real-time subtraction technique. The measured probe sensitivity for 18F was 1.1 cps kBq-1 ml-1 for the small head and 3.4 cps kBq-1 ml-1 for the large head. The mean spatial resolution was 1.6 mm FWHM on the detector surface. The γ-ray rejection efficiency measured by realistic brain phantom modelling of the surgical cavity was 99.4%. This phantom also demonstrated the ability of the probe to detect tumour discs as small as 5 mm in diameter (20 mg) for tumour-to-background ratios higher than 3:1 and with an acquisition time around 4 s at each scanning step. These results indicate that our detector could be a useful complement to existing techniques for the accurate excision of brain tumour tissue and more generally to improve the efficiency of radio-guided cancer surgery.
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/54/14/006Additional details
Identifiers
- DOI
- 10.1088/0031-9155/54/14/006;
- PII
- S0031-9155(09)11424-0;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 54
- Journal Issue
- 14
- Journal Page Range
- p. 4439-4453
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41050868
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BRAIN; GLIOMAS; POSITRONS; PROBES; SURGERY
- Descriptors DEC
- ANTILEPTONS; ANTIMATTER; ANTIPARTICLES; BODY; CENTRAL NERVOUS SYSTEM; DISEASES; ELEMENTARY PARTICLES; FERMIONS; LEPTONS; MATTER; MEDICINE; NEOPLASMS; NERVOUS SYSTEM; NERVOUS SYSTEM DISEASES; ORGANS