Published June 2008 | Version v1
Journal article

The Compton source efficiency tracing method in liquid scintillation counting: A new standardization method using a TDCR counter with a Compton spectrometer

  • 1. LNE-Laboratoire National Henri Becquerel, CEA-Saclay, F91191 Gif-sur-Yvette Cedex (France)

Description

We describe a new standardization method in liquid scintillation counting based on the use of a temporary virtual tracer source created inside the scintillator by Compton interaction. The Compton tracer source is measured by the triple coincidence liquid scintillation counter after selection of the Compton events by a gamma-ray detector. The paper describes the principle of the method, the experimental setup and presents the results obtained by this method for the standardization of a 3H solution. These results are compared with those obtained using the classical triple to double coincidence ratio method. Possible application of this new method for international comparisons is addressed

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2008.02.062

Additional details

Identifiers

DOI
10.1016/j.apradiso.2008.02.062;
PII
S0969-8043(08)00103-6;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
66
Journal Issue
6-7
Journal Page Range
p. 1026-1032
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
16. international conference on radionuclide metrology and its applications
Dates
3-7 Sep 2007
Place
Cape Town (South Africa)

Optional Information

Copyright
Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.