Attenuation and scatter correction in planar gamma camera imaging

  • 1. Faculty of Electrical Engineering, University of Belgrade, 11000 Belgrade (Serbia)

Description

A simple and a powerful method to correct for attenuation and the buildup of scatter in planar gamma camera nuclear medicine imaging is presented. The method is based on use of uniform μ-map to correct for attenuation and Monte Carlo simulated geometric progression dose buildup factors to correct for buildup of scatter. A comparison of results obtained by proposed method and Sorenson geometric mean approach is presented for some phantoms. Excellent results were obtained (errors of less than % 3 ) for various phantom configurations used to evaluate the method. (author)

Files

41123731.pdf

Files (314.1 kB)

Name Size Download all
md5:8c61be81902d21234f42128c3739c60f
314.1 kB Preview Download
Part of:
Proceedings of the 5th Yugoslav Nuclear Society Conference (YUNSC-2004)

Additional details

Publishing Information

Publisher
Institute of Nuclear Sciences VINCA
Imprint Place
Belgrade (Serbia)
ISBN
86-7306-076-1
Imprint Title
Proceedings of the 5th Yugoslav Nuclear Society Conference (YUNSC-2004)
Imprint Pagination
544 p.
Journal Page Range
p. 399-403
Report number
INIS-CS--0005

Conference

Title
5. International Yugoslav Nuclear Society Conference
Acronym
YUNSC-2004
Dates
27-30 Sep 2004
Place
Belgrade (Yugoslavia)

Optional Information

Notes
10 refs., 7 figs.; This record replaces 36115503