Full-field speckle interferometry for non-contact photoacoustic tomography
- 1. Medical Laser Center Lübeck GmbH, Peter-Monnik-Weg 4, 23562 Lübeck (Germany)
- 2. Institute of Biomedical Optics, University of Lübeck, Peter-Monnik-Weg 4, 23562 Lübeck (Germany)
Description
A full-field speckle interferometry method for non-contact and prospectively high speed Photoacoustic Tomography is introduced and evaluated as proof of concept. Thermoelastic pressure induced changes of the objects topography are acquired in a repetitive mode without any physical contact to the object. In order to obtain high acquisition speed, the object surface is illuminated by laser pulses and imaged onto a high speed camera chip. In a repetitive triple pulse mode, surface displacements can be acquired with nanometre sensitivity and an adjustable sampling rate of e.g. 20 MHz with a total acquisition time far below one second using kHz repetition rate lasers. Due to recurring interferometric referencing, the method is insensitive to thermal drift of the object due to previous pulses or other motion. The size of the investigated area and the spatial and temporal resolution of the detection are scalable. In this study, the approach is validated by measuring a silicone phantom and a porcine skin phantom with embedded silicone absorbers. The reconstruction of the absorbers is presented in 2D and 3D. The sensitivity of the measurement with respect to the photoacoustic detection is discussed. Potentially, Photoacoustic Imaging can be brought a step closer towards non-anaesthetized in vivo imaging and new medical applications not allowing acoustic contact, such as neurosurgical monitoring or burnt skin investigation. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/60/10/4045Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 60
- Journal Issue
- 10
- Journal Page Range
- p. 4045-4058
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47072558
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; IMAGES; IN VIVO; INTERFEROMETRY; PHANTOMS; SENSITIVITY; SILICONES; SKIN; TOMOGRAPHY
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANIC COMPOUNDS; ORGANIC SILICON COMPOUNDS; ORGANS; POLYMERS; RADIOLOGY; SILOXANES; STRUCTURAL MODELS