Optimisation of a secondary standard chamber for the measurement of the ambient dose equivalent, H*(10), for low photon energies

Creators

  • 1. Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig (Germany)

Description

A secondary standard ionisation chamber for photon radiation for measuring an ionisation current, which is directly proportional to the conventionally true value of the ambient dose equivalent, H* (10), was optimised. The chamber was developed in the Austrian Research Centers Seibersdorf and is used successfully worldwide by dosimetry laboratories. The chamber response with respect to H*(10) for photon energies from 40 to 1250 keV is nearly constant. For lower photon energies the response is strongly energy-dependent and does not fulfil the requirements concerning the quality of a secondary standard given in ISO 4037-2, i.e. for energies for which the determination of the conventionally true value of H*(10) is very difficult. Considering the dose limits defined in the Directive 96/29/Euratom, in the case of whole-body irradiation the knowledge of the personal dose equivalent is of importance down to energies of ∼12 keV. For area dosimetry, this means that the knowledge of H*(10) for energies approximately ≥12 keV is necessary. To get one secondary standard chamber for H*(10) for the whole photon energy range and to close the gap for low energies in the dissemination of the conventionally true value of H*(10), the chamber was optimised for a flat response for energies from ∼12 to 1250 keV. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nci334

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
118
Journal Issue
1
Journal Page Range
p. 16-21
ISSN
0144-8420

Optional Information

Notes
15 refs