Optical constants of magnetic spark spectrometer
- 1. Joint Inst. for Nuclear Research, Dubna (USSR). Lab. of Nuclear Problems
- 2. Helsinki Univ. (Finland). Dept. of High Energy Physics
- 3. Milan Univ. (Italy). Ist. di Fisica
Description
For geometrical reconstruction of events, registered on two films, in the space of magnetic spark spectrometer the mathematical model of the spectrometer optics is constructed. It couples measurements of fiducial marks on film with space position in order to find parameters of the system which are the optical constants. The search of spectrometer optical constants is an iterative process during which the parameters in question are found in a modified model taking into account the distortion of measurements and distortion in space due to stretching of film, shifting marks in magnetic field and optical distortion. The optical constants found allow one to achieve high accuracy reconstruction of fiducial marks and particle tracks and to determine the momentum and angular resolution
Availability note (English)
MF available from INIS under the Report Number.Files
14743114.pdf
Files
(199.4 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:a155bd6b11b87b30809f5d345fdd5b9e
|
199.4 kB | Preview Download |
System files
(31.1 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Оптические константы спектрометра MIS
Publishing Information
- Imprint Pagination
- 7 p.
- Report number
- JINR--10-82-227
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 14743114
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- LINEAR MOMENTUM RESOLUTION; MAGNETIC SPECTROMETERS; NEWTON METHOD; OPTICAL SYSTEMS; PARTICLE TRACKS
- Descriptors DEC
- ITERATIVE METHODS; MEASURING INSTRUMENTS; RESOLUTION; SPECTROMETERS
Optional Information
- Notes
- 3 refs.; 1 fig.