Published August 2023 | Version v1
Journal article

Experimentally determined relative biological effectiveness of cyclotron-based epithermal neutrons designed for clinical BNCT. In vitro study

  • 1. Kyoto University, Institute for Integrated Radiation and Nuclear Science, Particle Radiation Oncology Research Center, Industrial Equipment Division, Kumatori, Osaka (Japan)
  • 2. Osaka Metropolitan University, Research Center for Boron Neutron Capture Therapy (BNCT), Sakai, Osaka (Japan)

Description

A neutron beam for boron neutron capture therapy (BNCT) of deep-seated tumours is designed to maintain a high flux of epithermal neutrons, while keeping the thermal and fast neutron component as low as possible. These neutrons (thermal and fast) have a high relative biological effectiveness in comparison with high energy photon beams used for conventional X-ray radiotherapy. In the past, neutrons for the purpose of BNCT were generated using nuclear reactors. However, there are various challenges that arise when installing a reactor in a hospital environment. From 2006, the Kyoto University Research Reactor Institute, in collaboration with Sumitomo Heavy Industries, began the development of an accelerator-based neutron source for clinical BNCT in a bid to overcome the shortcomings of a nuclear reactor-based neutron source. Following installation and beam performance testing, in vitro studies were performed to assess the biological effect of the neutron beam. Four different cell lines were prepared and irradiated using the accelerator-based neutron source. Following neutron and gamma ray irradiation, the survival curve for each cell line was calculated. The biological end point to determine the relative biological effectiveness (RBE) was set to 10% cell survival, and the D10 for each cell line was determined. The RBE of the accelerator-based neutron beam was evaluated to be 2.62. (author)

Availability note (English)

Available from DOI: https://doi.org/10.1093/jrr/rrad056

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Radiation Research (Online)
Journal Volume
64
Journal Issue
5
Journal Page Range
p. 811-815
ISSN
1349-9157

Optional Information

Notes
14 refs., 6 figs., 2 tabs.