Published April 1989 | Version v1
Report Open

An alignment method for the Mark II silicon strip vertex detector using an x-ray beam

Description

A silicon strip vertex detector consisting of 36 independent detector modules is being constructed for use in the Mark II detector at the SLAC Linear Collider. This paper describes a method for determining the relative alignment of the modules to a precision better than the 5 μm intrinsic resolution of the detectors. The basic procedure involves moving the vertex detector by known amounts through a fixed, collimated x-ray beam, and using the beam position reconstructed from the detected signals to determine the relative positions and orientations of the modules. Results from tests of the method on a subset of detectors are presented. 5 refs., 11 figs

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A03/MF A01 - OSTI; 1 as DE89014588.

Files

20072229.pdf

Files (454.1 kB)

Name Size Download all
md5:b1f2de59d4e7a015fc737382080186bf
454.1 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
12 p.
Report number
SLAC-PUB--4961

Conference

Title
new developments on radiation detectors.
Acronym
5. European symposium on semiconductor detectors
Dates
21-23 Feb 1989.
Place
Munich (Germany, F.R.).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
20072229
Subject category
S43: PARTICLE ACCELERATORS; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALIGNMENT; SHOWER COUNTERS; SI SEMICONDUCTOR DETECTORS; STANFORD LINEAR COLLIDER; TESTING; X-RAY SOURCES
Descriptors DEC
ACCELERATORS; LINEAR ACCELERATORS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATION SOURCES; SEMICONDUCTOR DETECTORS

Optional Information

Notes
Portions of this document are illegible in microfiche products.
Secondary number(s)
CONF-890247--3.