Published August 27, 2003
| Version v1
Report
Analysis of Neutron and Photon Detection Position for the Calibration of Plastic (BC-420) and Liquid (BC-501) Scintillators
Creators
Description
This report presents the results of the neutron and photon calibration of a Bicron BC-420 plastic scintillator and a BC-501 liquid scintillator using a set of reference gamma sources and a Cf-252 source. The position of neutron and photon detection inside the scintillators as a function of energy is investigated by Monte Carlo simulation
Availability note (English)
Available from http://www.ornl.gov/~webworks/cppr/y2001/rpt/117734.pdf; PURL: https://www.osti.gov/servlets/purl/885588-iS4qWb/Additional details
Identifiers
Publishing Information
- Imprint Pagination
- 23 p.
- Report number
- ORNL/TM--2003/168
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 37104899
- Subject category
- S72: PHYSICS OF ELEMENTARY PARTICLES AND FIELDS; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Non-conventional Literature
- Descriptors DEI
- CALIBRATION; COMPUTERIZED SIMULATION; GAMMA DETECTION; GAMMA SOURCES; LIQUID SCINTILLATORS; NEUTRON DETECTION; NEUTRONS; PHOSPHORS; PHOTONS; PLASTIC SCINTILLATORS
- Descriptors DEC
- BARYONS; BOSONS; DETECTION; ELEMENTARY PARTICLES; FERMIONS; HADRONS; MASSLESS PARTICLES; NUCLEONS; PHOSPHORS; RADIATION DETECTION; RADIATION SOURCES; SIMULATION
Optional Information
- Contract/Grant/Project number
- AC05-00OR22725
- Funding organization
- US Department of Energy (United States)