Oversimplified, compact double focusing mass spectrometer geometries

  • 1. Institute for Isotopic and Molecular Technology, Cluj (Romania)

Description

The simplest ways to construct double focusing mass spectrometer designs are those resulting from inserting plane condenser plates in the gap of two rectangular pole pieces or a cylindric condenser at the pole piece corner. The first order ion-optical parameters and aberration coefficients were calculated for double focusing Wien filter-magnetic deflector geometries over the useful range of the deflection angles. Similar calculations were performed for cylindric electric-magnetic crossed field sectors, for comparison. The illustrating numerical examples indicate the highest resolutions for the 90 deflecting Wien filter-magnet and crossed field geometries. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
363
Journal Issue
1-2
Journal Page Range
p. 406-410.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
4. international conference on charged particle optics (CPO-4).
Dates
3-6 Oct 1994.
Place
Tsukuba, Ibaraki (Japan).