Overview of equilibrium reconstruction on DIII-D using new measurements from an expanded motional Stark effect diagnostic
- 1. Lawrence Livermore National Laboratory, 7000 East Ave., Livermore, California 94550 (United States)
- 2. General Atomics, P.O. Box 85608, San Diego, California 92186-5608 (United States)
Description
Motional Stark effect (MSE) measurements constrain equilibrium reconstruction of DIII-D tokamak plasmas using the equilibrium code EFIT. In 2007, two new MSE arrays were brought online, bringing the system to three core arrays, two edge arrays, and 64 total channels. We present the first EFIT reconstructions using this expanded system. Safety factor and ER profiles produced by fitting to data from the two new arrays and one of the other three agree well with independent measurements. Comparison of the data from the three arrays that view the core shows that one of the older arrays is inconsistent with the other two unless the measured calibration factors for this array are adjusted. The required adjustments depend on the toroidal field and plasma current direction, and on still other uncertain factors that change as the plasma evolves. We discuss possible sources of calibration error for this array.
Additional details
Identifiers
- DOI
- 10.1063/1.2955711;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 79
- Journal Issue
- 10
- Journal Page Range
- p. 10F518-10F518.4
- ISSN
- 0034-6748
- CODEN
- RSINAK
Conference
- Title
- 17. topical conference on high-temperature plasma diagnostics
- Acronym
- HTPD08
- Dates
- 11-15 May 2008
- Place
- Albuquerque, NM (United States)
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 41016162
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; COMPARATIVE EVALUATIONS; DOUBLET-3 DEVICE; ELECTRIC CURRENTS; EQUILIBRIUM; PLASMA; PLASMA DIAGNOSTICS; STARK EFFECT
- Descriptors DEC
- CLOSED PLASMA DEVICES; CURRENTS; EVALUATION; THERMONUCLEAR DEVICES; TOKAMAK DEVICES
Optional Information
- Notes
- (c) 2008 American Institute of Physics