Published June 2003 | Version v1
Report Open

Monte-Carlo simulation of the SL-ELEKTA-20 medical linear accelerator. Dosimetric study of a water phantom

Description

In radiotherapy, it is essential to have a precise knowledge of the dose delivered in the target volume and the neighbouring critical organs. To be usable clinically, the models of calculation must take into account the exact characteristics of the beams used and the densities of fabrics. Today we can use sophisticated irradiation techniques and get a more precise assessment of the dose and with a better knowledge of its distribution. Thus in this report, will be detailed a simulation of the head of irradiation of accelerator SL-ELEKTA-20 in electrons mode and a dosimetric study of a water phantom. This study is carried out with the code of simulation Monte Carlo GATE adapted for applications of medical physics; the results are compared with the data obtained by the anticancer center 'Jean Perrin' on a similar accelerator. (author)

Availability note (English)

Available from INIS in electronic form

Files

37017867.pdf

Files (1.3 MB)

Name Size Download all
md5:009772c99fa80d1d993c0dd992dfe6a0
1.3 MB Preview Download

Additional details

Additional titles

Original title (French)
Simulation Monte Carlo de l'accelerateur lineaire clinique SL-ELEKTA 20. Etude dosimetrique dans un fantome d'eau

Publishing Information

Imprint Pagination
42 p.
Report number
FRNC-TH--6187

Optional Information

Notes
8 refs