Published December 1, 1985
| Version v1
Journal article
Multichannel flash x-ray tomography of microsecond events
Description
The design and operational characteristics of the Ballistic Research Laboratory tomography facility for the study of density profiles of rapidly moving objects are described. It is shown that time resolution in the microsecond range and spatial resolution of the order of millimeters with density resolution of the order of 1% are feasible even when the x-ray transmission path contains several centimeters of high atomic number materials
Additional details
Publishing Information
- Journal Title
- Appl. Opt.
- Journal Volume
- 24
- Journal Issue
- 23
- Series
- Appl. Opt.
- Journal Page Range
- 4061-4063
- ISSN
- 0003-6935
- CODEN
- APOPA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18062012
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE; S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; DYNAMIC FUNCTION STUDIES; IMAGE SCANNERS; INDUSTRIAL RADIOGRAPHY; PERFORMANCE; SPATIAL RESOLUTION; SPECIFICATIONS; TIME RESOLUTION; X RADIATION
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MATERIALS TESTING; NONDESTRUCTIVE TESTING; RADIATIONS; RESOLUTION; TESTING; TIMING PROPERTIES; TOMOGRAPHY