Published December 2006 | Version v1
Journal article

Experimental phantom lesion detectability study using a digital breast tomosynthesis prototype system

  • 1. Radiologisches Inst., Univ. Erlangen-Nuernberg (Germany)
  • 2. Siemens AG Medical Solutions, Technik und Medizinphysik (Germany)

Description

Purpose: To compare the sensitivity of conventional two-dimensional (2D) projection imaging with tomosynthesis with respect to the detectability of mammographic phantom lesions. Materials and Methods: Using a breast tomosynthesis prototype based on a commercial FFDM system (Siemens MAMMOMAT NovationDR), but modified for a wide angle tube motion and equipped with a fast read-out amorphous selenium detector, we acquired standard 2D images and tomosynthesis series of projection views. We used the Wisconsin mammographic random phantom, model RMI 152A. The anode filter combinations Mo/Mo and W/Rh at two different doses were used as typical radiographic techniques. Slice images through the phantom parallel to the detector were reconstructed with a distance of 1 mm employing a filtered back-projection algorithm. The image data sets were read by five radiologists and evaluated with respect to the detectability of the phantom details. Results: For all studied radiographic techniques, the detection rate in the tomosynthesis mode was 100%, i.e. 75 true positive findings out of 75 possible hits. In contrast, the conventional projection mode yielded a detection rate between 80 and 93% (corresponding to 60 and 70 detected details) depending on the dose and X-ray spectrum. Conclusion: Tomosynthesis has the potential to increase the sensitivity of digital mammography. Overlapping structures from out-of-plane tissue can be removed in the tomosynthesis reconstruction process, thereby enhancing the diagnostic accuracy. (orig.)

Additional details

Publishing Information

Journal Title
RoeFo - Fortschritte auf dem Gebiete der Roentgenstrahlen und der Bildgebenden Verfahren
Journal Volume
178
Journal Issue
12
Journal Page Range
p. 1219-1223
ISSN
1438-9029
CODEN
RFGNDO

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
38039991
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; DIGITAL SYSTEMS; MAMMARY GLANDS; NEOPLASMS; PHANTOMS
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DISEASES; EVALUATION; GLANDS; MOCKUP; ORGANS; STRUCTURAL MODELS; TOMOGRAPHY