Published 1981 | Version v1
Journal article

Automatic stabilization of relative counting losses in gamma-ray spectrometry of short-lived nuclides

  • 1. Institute of Nuclear Research, Warsaw (Poland)

Description

An electronic circiut for the stabilization of the relative counting losses due to dead time and pile-up effects is described. The circuit consists of two independent channels: for stabilization of dead time and pile-up, respectively. The stabilizer (circuit) receives continuously information on temporary dead time and pile-up in a spectrometer and owing to feedback the relative counting losses (in peaks) are constant during the measuring time and can be easily calculated. (author)

Additional details

Publishing Information

Journal Title
J. Radioanal. Chem.
Journal Volume
61
Journal Issue
1-2
Series
J. Radioanal. Chem.
Journal Page Range
93-110
ISSN
0134-0719

Conference

Title
Workshop on activation analysis with short-lived nuclides.
Dates
4 - 8 Feb 1980.
Place
Vienna, Austria.

Optional Information

Notes
33 refs.