Published 1981
| Version v1
Journal article
Automatic stabilization of relative counting losses in gamma-ray spectrometry of short-lived nuclides
Description
An electronic circiut for the stabilization of the relative counting losses due to dead time and pile-up effects is described. The circuit consists of two independent channels: for stabilization of dead time and pile-up, respectively. The stabilizer (circuit) receives continuously information on temporary dead time and pile-up in a spectrometer and owing to feedback the relative counting losses (in peaks) are constant during the measuring time and can be easily calculated. (author)
Additional details
Publishing Information
- Journal Title
- J. Radioanal. Chem.
- Journal Volume
- 61
- Journal Issue
- 1-2
- Series
- J. Radioanal. Chem.
- Journal Page Range
- 93-110
- ISSN
- 0134-0719
Conference
- Title
- Workshop on activation analysis with short-lived nuclides.
- Dates
- 4 - 8 Feb 1980.
- Place
- Vienna, Austria.
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 12605655
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AUTOMATION; CORRECTIONS; DEAD TIME; ELECTRONIC CIRCUITS; ERRORS; FLOWSHEETS; GAMMA SPECTROSCOPY; LI-DRIFTED GE DETECTORS; LOSSES; MULTI-CHANNEL ANALYZERS; RESOLUTION; STABILITY
- Descriptors DEC
- DIAGRAMS; ELECTRONIC EQUIPMENT; GE SEMICONDUCTOR DETECTORS; INFORMATION; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; PULSE ANALYZERS; RADIATION DETECTORS; SEMICONDUCTOR DETECTORS; SPECTROSCOPY; TIMING PROPERTIES
Optional Information
- Notes
- 33 refs.