SU-F-J-53: A 3D Printed Phantom for the Use of Daily Quality Assurance Alignment Tests
Creators
- 1. Ohio State University Columbus, OH (United States)
Description
Purpose: To discuss experiences and results for a 3D printed QA phantom used for daily alignment purposes for a six degrees-of-freedom (6DoF) table Methods: A 3D model was created using a fused deposition modeling (FDM) printer using free online computer-aided design (CAD) software. The model has been under use for daily QA alignment tests for a 6DoF couch. An aligned and angled baseplate were also printed in order to introduce known angles for 6DoF corrections during image-guidance. Unique registration contours were created on the faces of the phantom in order to achieve a better cone-beam computed tomography (CBCT) match using an auto-registration algorithm. A BB was also introduced at the center of the phantom in order to deliver an integrated daily Winston-Lutz (WL) test. Translational, rotational, and WL results were tabulated over one month. Results: The 'honeycomb' structure of the print was apparent in the EPID images for the WL test, which affected the results of the analysis software. This was fixed by inserting a cube made of polyoxymethylene within the 3D phantom that encompass the BB. Auto-registration results for the three translational and three rotational from a known offset to the BB isocenter consistently fell within 1 mm and 0.2°, respectively. WL tests resulted in an average of 0.71 ± 0.14 mm. Conclusion: 3D printed models allow for accurate builds that can be customized to a variety of clinical needs. Results from translational, rotational, and WL show consistent results over a month's time. Given its relatively cheap and streamlined workflow, 3D printing could be implemented into any clinic looking to create customized phantoms.
Additional details
Identifiers
- DOI
- 10.1118/1.4955961;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 43
- Journal Issue
- 6
- Journal Page Range
- vp.
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 48080351
- Subject category
- S60: APPLIED LIFE SCIENCES; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ALGORITHMS; COMPUTER CODES; COMPUTER-AIDED DESIGN; COMPUTERIZED TOMOGRAPHY; CORRECTIONS; HONEYCOMB STRUCTURES; IMAGES; PHANTOMS; POLYOXYMETHYLENES; QUALITY ASSURANCE
- Descriptors DEC
- DESIGN; DIAGNOSTIC TECHNIQUES; MATHEMATICAL LOGIC; MECHANICAL STRUCTURES; MOCKUP; ORGANIC COMPOUNDS; ORGANIC POLYMERS; POLYACETALS; POLYMERS; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- (c) 2016 American Association of Physicists in Medicine