Published March 11, 2004 | Version v1
Journal article

Demonstration of feasibility of operating a silicon ZIP detector with 20 eV threshold

Description

100-g silicon detectors (known as 'ZIPs', Z-resolving Ionization and Phonon detectors) developed by the Cryogenic Dark Matter Search Experiment have been tested at charge bias voltages of up to 200 V/cm, significantly above their usual operating range (∼3-6 V/cm). Thermal gain factors in excess of 50 were observed due to the primary ionization drifting in the large applied field, with only minimal increase in phonon noise. The observed thermal gain corresponds to an intrinsic threshold of 20 eV, resulting in detectors that have direct application for use in a neutrino magnetic moment measurement based on a 40-MCi tritium source

Additional details

Identifiers

DOI
10.1016/j.nima.2003.11.283;
PII
S0168900203031917;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
520
Journal Issue
1-3
Journal Page Range
p. 163-166
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
10. international workshop on low temperature detectors
Dates
7-11 Jul 2003
Place
Genoa (Italy)

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.