Published March 2002 | Version v1
Journal article

Sensitometry of intensifying screen-film system for detection of yearly change of imaging quality factors

  • 1. Hiroshima Prefectural Coll. of Health Sciences, Mihara (Japan)
  • 2. Kagawa Medical Univ., Miki (Japan). Hospital
  • 3. Hiroshima Univ. (Japan)

Description

Systematic sensitometry of 20 green- and 25 blue-intensifying screens made in early 1980s and early 1990s was performed on their physical imaging qualities to evaluate their yearly change. Sensitometry was done with Toshiba KXO-1000 X-ray generator (Tube: Toshiba DRX-431HD, 80 kV, 20 mm-thick Al-filter), characteristic curve examination with KXO-15 (Tube: DRX-1603B, 80 kV, 0.5 mm Cu+4.0 mm Al), determination of modulation transfer function with KXO-15 (Tube: DRX-1303B, 80 kV, 2.0 mm Al), and Wiener spectrometry with Hitachi DH-158 (Tube: Hitachi UH-6GC-108DL, 80 kV, 20 mmAl). Reference screen was that reported by US Bureau of Radiological Health (1982) and Center for Devices and Radiological Health (196) and was obtained from Chicago University. There were products without yearly change, with increased sensitivity with unchanged resolution and noise level, with improved resolution, or with improved sensitivity. (K.H.)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
58
Journal Issue
3
Journal Page Range
p. 322-329
ISSN
0369-4305

Optional Information