Description

This paper presents the performance evaluation of radionuclide calibrators of Nuclear Medicine Centers located in Rio de Janeiro state and Porto Alegre city at the criterion of accuracy required by the standard NN-3.05 of the National Commission on Nuclear Energy for measuring of activity of radiopharmaceuticals. Of total of 203 results evaluated for 99mTc, 131I, 67Ga and 201Tl, 88% showed acceptable performance according to this criterion. Ideally, and fully attainable, the performance should be 100%, for the benefit of patients undergoing nuclear medicine procedures. (author)