Acceptance testing and routine quality control in general radiography: Mobile units and film/screen fixed systems
Creators
- 1. St. James's Hospital, The Haughton Inst., Dublin (Ireland)
Description
This study presents the findings from acceptance testing and routine quality control (QC) of general radiographic X-ray equipment in Ireland during 2006 and early 2007, including mobile X-ray units and film/screen fixed systems. Acceptance testing and routine QC of the diagnostic X-ray imaging equipment are requirements of European and Irish legislation. One hundred general radiographic X-ray systems were tested within Ireland, 73% of them failed to meet the required QC guidelines, whereby one or more faults were identified. The majority of these failures were minor ones, requiring attention by the suppliers at the next routine service. Significant faults were only identified in seven systems. The suppliers were requested to investigate these issues as soon as possible and take the necessary corrective action. A review of the QC results highlights the need to perform comprehensive acceptance and routine testing of the systems. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncn033Additional details
Identifiers
- DOI
- 10.1093/rpd/ncn033;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 129
- Journal Issue
- 1-3
- Journal Page Range
- p. 276-278
- ISSN
- 0144-8420
Conference
- Title
- EC SENTINEL Project workshop. Safety and efficacy for new techniques and imaging using new equipment to support European legislation
- Dates
- 18-20 Apr 2007
- Place
- Delf (Netherlands)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 43120781
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- FILMS; HUMAN POPULATIONS; IMAGES; QUALITY CONTROL; TESTING; X RADIATION; X-RAY EQUIPMENT
- Descriptors DEC
- CONTROL; ELECTROMAGNETIC RADIATION; EQUIPMENT; IONIZING RADIATIONS; POPULATIONS; RADIATIONS
Optional Information
- Notes
- 7 refs