Evaluation of different 16-row CT colonography protocols using a porcine model

  • 1. Tuebingen Univ. (Germany). Abt. fuer Radiologische Diagnostik
  • 2. Universitaet Tuebingen (Germany). Institut fuer Biometrie

Description

Purpose: To develop and to test an easily produced biological colon model with simulated polypoid lesions. Application of this phantom for the selection of an optimized scan protocol of 16-row CT colonography (CTC) for clinical use. Methods and Material: Six polypoid lesions (1-6 mm) were simulated with sutures on the inner face of a porcine colon segment (20 cm). After distending the colon segment with air, the phantom was placed in a water quench and CT scans were performed on a MDCT-scanner (Somatom Sensation 16, Siemens, Forchheim). At constant values for collimation (16 x 0.75 mm) and voltage (120 kV), 54 different combinations of mAs values (50, 75 and 100 mAs), pitch factors (1, 1.25 and 1.5) and slice thicknesses (0.75, 1.0, 1.5, 2.0, 3.0 and 5.0 mm) were tested systematically. The phantom was scanned in the longitudinal and transverse axis to simulate the different orientation of the colon in the abdomen. Axial slice images and virtual endoscopic views of all data sets were presented separately to 2 radiologists who independently determined number and size of detectable polyps. Dose exposure was measured with an Alderson phantom. Results: The colon model offered a realistic imitation of a polyp-covered, human colon. The experimental set-up allowed a systematic evaluation of polyp detection related to lesion size, orientation of the colon and CTC parameters, with other influencing factors mostly excluded. Polyps were significantly better detected in the longitudinal than in the transverse orientation of the colon. For the detection of lesions of at least 3 mm, a low dose (50 mAs) 16-row CTC should be combined with a pitch of 1.5 and a maximum slice thickness of 3 mm. For the depiction of polyps smaller than 3 mm, slice thickness and pitch should amount to 1 mm and 1.0, respectively. Effective dose of this low dose protocol is 4.08 mSv. (orig.)

Additional details

Additional titles

Original title (German)
Evaluation unterschiedlicher 16-Zeilen-CT-Colonographie-Protokolle am Schweinedarmmodell

Publishing Information

Journal Title
RoeFo - Fortschritte auf dem Gebiete der Roentgenstrahlen und der bildgebenden Verfahren
Journal Volume
176
Journal Issue
10
Journal Page Range
p. 1493-1500
ISSN
1438-9029
CODEN
RFGNDO