A new neutron monitor and extended conversion coefficients for HP(10)
Creators
- 1. Dipartimento di Ingegneria Meccanica, Nucleare e della Produzione, I-56126 Pisa (Italy)
- 2. Physikalisch-Technische Bundesanstalt PTB, D-38116 Braunschweig (Germany)
Description
A new personal dose equivalent monitor for neutrons, the 'HpSLAB', is introduced. The device consists of a 30 x 30 x 15 cm3 polymethyl-methacrylate slab hosting a superheated drop detector embedded at a depth of 10 mm. The personal dose equivalent monitor was characterised experimentally with fast neutron calibrations in the 0.144-14.8 MeV range and numerically with Monte Carlo simulations. In order to evaluate the performance of the device, its response was compared to the fluence-to-directional dose equivalent conversion coefficients, hp(10;α,E). Since published coefficients only cover neutron angles of incidence up to 75 deg., a new extended set of coefficients was computed for angles of incidence up to 180 deg.. The method used in these calculations was the very same used in the generation of the dose equivalent coefficients recommended by International Commission on Radiological Protection publication 74. The response of the HpSLAB follows with good approximation the trend of the conversion coefficients for monoenergetic neutrons above ∼5 MeV. The device was extensively tested in broad-spectrum workplace-fields encountered at nuclear installations and its response was on average within a factor 1.4 of the reference personal dose equivalent values, regardless of angle and energy distribution of the neutron fluence. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncm316Additional details
Identifiers
- DOI
- 10.1093/rpd/ncm316;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 125
- Journal Issue
- 1-4
- Journal Page Range
- p. 345-348
- ISSN
- 0144-8420
Conference
- Title
- European Workshop on Individual Monitoring of Ionising Radiation
- Acronym
- IM2005
- Dates
- 11-15 Apr 2005
- Place
- Vienna (Austria)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44022226
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; COMPUTERIZED SIMULATION; DOSE EQUIVALENTS; ENERGY SPECTRA; ENVIRONMENTAL EXPOSURE; FAST NEUTRONS; INCIDENCE ANGLE; KEV RANGE 100-1000; METHACRYLATES; MEV RANGE 01-10; MEV RANGE 10-100; MONTE CARLO METHOD; NEUTRON FLUENCE; NEUTRON MONITORS; NUCLEAR FACILITIES; PERFORMANCE; RADIATION MONITORING; RADIATION PROTECTION
- Descriptors DEC
- BARYONS; CALCULATION METHODS; CARBOXYLIC ACID SALTS; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; HADRONS; KEV RANGE; MEASURING INSTRUMENTS; MEV RANGE; MONITORING; MONITORS; NEUTRONS; NUCLEONS; RADIATION MONITORS; SIMULATION; SPECTRA
Optional Information
- Notes
- 16 refs