Compressive sensing for polarization sensitive optical coherence tomography
Creators
- 1. Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, IL 61801 (United States)
Description
In this report, we report on the implementation of compressive sensing (CS) and sparse sampling in polarization sensitive optical coherence tomography (PS-OCT) to reduce the number of B-scans (frames consisting of an array of A-scans, where each represents a single depth profile of reflections) required for effective volumetric (3D dataset composed of an array of B-scans) PS-OCT measurements (i.e. OCT intensity, and phase retardation) reconstruction. Sparse sampling of PS-OCT is achieved through randomization of step sizes along the slow-axis of PS-OCT imaging, covering the same spatial ranges as those with equal slow-axis step sizes, but with a reduced number of B-scans. Tested on missing B-scan rates of 25%, 50% and 75%, we found CS could reconstruct reasonably good (as evidenced by a correlation coefficient >0.6) PS-OCT measurements with a maximum reduced B-scan rate of 50%, thereby accelerating and doubling the rate of volumetric PS-OCT measurements. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6463/abf958Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. D, Applied Physics
- Journal Volume
- 54
- Journal Issue
- 29
- Journal Page Range
- [6 p.]
- ISSN
- 0022-3727
- CODEN
- JPAPBE
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 53060741
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
- Descriptors DEI
- POLARIZATION; SAMPLING; TOMOGRAPHY
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES