Published January 2013 | Version v1
Journal article

Determination of monitor unit check tolerances based on a comparison with measurement and treatment planning system data

  • 1. Medical Physics Department, The James Cook University Hospital, Middlesbrough (United Kingdom)

Description

ABSTRACT: This work describes the experimental validation of treatment planning system monitor unit (MU) calculations against measurement for a range of scenarios. This, together with a comparison of treatment planning system MUs and an independent MU check method, allows the derivation of confidence intervals for the check process. Data were collected for open and 60° motorized wedge fields using an Elekta Synergy linac at 6 and 8 MV using homogeneous and heterogeneous phantoms. Masterplan (Version 4.0) pencil-beam and collapsed cone algorithms were used for the primary MU calculations with full inhomogeneity correction. Results show that both algorithms agree with measurement to acceptable tolerance levels in the majority of the cases studied. The confidence interval for the pencil-beam algorithm MU against an independent check was determined as + 1.6% to −3.4%. This is modified to + 2.3% to −2.5% when data collected with low-density heterogeneities are removed as this algorithm is not used clinically for these cases. The corresponding interval for the collapsed cone algorithm was + 1.2% to −4.3%, indicating that an offset tolerance for the independent check is appropriate. Analysis of clinical conformal treatment plan data generated using the pencil-beam algorithm (1393 beams) returned 93% of beams within the independent check tolerance. Similarly, using the collapsed cone algorithm as the primary MU calculation, 77% (of 1434 beams) were within the confidence interval

Availability note (English)

Available from http://dx.doi.org/10.1016/j.meddos.2012.07.005

Additional details

Identifiers

DOI
10.1016/j.meddos.2012.07.005;
PII
S0958-3947(12)00138-0;

Publishing Information

Journal Title
Medical Dosimetry
Journal Volume
38
Journal Issue
1
Journal Page Range
p. 81-87
ISSN
0958-3947

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
45084022
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
ALGORITHMS; COMPARATIVE EVALUATIONS; EXTERNAL BEAM RADIATION THERAPY; LINEAR ACCELERATORS; PHANTOMS; PLANNING
Descriptors DEC
ACCELERATORS; EVALUATION; MATHEMATICAL LOGIC; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; RADIOTHERAPY; STRUCTURAL MODELS; THERAPY

Optional Information

Copyright
Copyright (c) 2012 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.