Published March 11, 2004 | Version v1
Journal article

Characterization of an Al-STJ-based X-ray detector with monochromatized synchrotron radiation

Description

The characterization of an Al-STJ-based detector with Pb absorber was performed with monochromatized synchrotron radiation. Detector response was measured in the energy range from 3 to 10 keV. A small non-linearity of the signal pulse height was detected, probably due to the escape of recombination phonons from the detector. The non-linearity can be described by a second order polynomial function. Additionally, detector signals were recorded while an X-ray beam of 50 μm diameter was directed to several locations on and near the absorber. For a well-aligned beam, detector artefacts are of at least two orders of magnitude lower intensity than the absorber events

Additional details

Identifiers

DOI
10.1016/j.nima.2003.11.353;
PII
S0168900203032066;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
520
Journal Issue
1-3
Journal Page Range
p. 234-236
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
10. international workshop on low temperature detectors
Dates
7-11 Jul 2003
Place
Genoa (Italy)

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35081015
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
KEV RANGE 01-10; PHONONS; POLYNOMIALS; PULSES; RADIATION DETECTION; RADIATION DETECTORS; RECOMBINATION; SYNCHROTRON RADIATION; X RADIATION
Descriptors DEC
BREMSSTRAHLUNG; DETECTION; ELECTROMAGNETIC RADIATION; ENERGY RANGE; FUNCTIONS; IONIZING RADIATIONS; KEV RANGE; MEASURING INSTRUMENTS; QUASI PARTICLES; RADIATIONS

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.