Published December 2021 | Version v1
Book

Performance test of Clover detector and BEGe detector

  • 1. Saha Institute of Nuclear Physics, Kolkata (India)

Description

Absolute γ-ray efficiency curves were obtained for an unsuppressed Clover detector and a Broad Energy Germanium (BEGe) detector in the energy range Eγ ∼ 50 - 1400 keV, by placing radioactive sources at 10 cm and 1 cm from face of the detectors. It was found that the absolute efficiency of an unsuppressed clover detector is considerably higher than the BEGe detector. The summing effect for the unsuppressed clover detector in a close geometry setup is much less than that for a BEGe detector. (author)

Part of:
Proceedings of the DAE-BRNS symposium on nuclear physics. V. 65

Additional details

Publishing Information

Publisher
Bhabha Atomic Research Centre
Imprint Place
Mumbai (India)
Imprint Title
Proceedings of the DAE-BRNS symposium on nuclear physics. V. 65
Imprint Pagination
[977 p.]
Journal Page Range
[2 p.]

Conference

Title
65. DAE-BRNS symposium on nuclear physics
Dates
1-5 Dec 2021
Place
Mumbai (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
53031795
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
GAMMA RADIATION; GE SEMICONDUCTOR DETECTORS; INSPECTION; PERFORMANCE TESTING; QUALITY CONTROL
Descriptors DEC
CONTROL; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; SEMICONDUCTOR DETECTORS; TESTING

Optional Information

Notes
Article No. G6