Imaging and quantification of anomaly volume using an eight-electrode 'hemiarray' EIT reconstruction method
Creators
- 1. J Crayton Pruitt Department of Biomedical Engineering, University of Florida, Box 116131, Gainesville, FL (United States)
Description
Electrical impedance tomography (EIT) is particularly well-suited to applications where its portability, rapid acquisition speed and sensitivity give it a practical advantage over other monitoring or imaging systems. An EIT system's patient interface can potentially be adapted to match the target environment, and thereby increase its utility. It may thus be appropriate to use different electrode positions from those conventionally used in EIT in these cases. One application that may require this is the use of EIT on emergency medicine patients; in particular those who have suffered blunt abdominal trauma. In patients who have suffered major trauma, it is desirable to minimize the risk of spinal cord injury by avoiding lifting them. To adapt EIT to this requirement, we devised and evaluated a new electrode topology (the 'hemiarray') which comprises a set of eight electrodes placed only on the subject's anterior surface. Images were obtained using a two-dimensional sensitivity matrix and weighted singular value decomposition reconstruction. The hemiarray method's ability to quantify bleeding was evaluated by comparing its performance with conventional 2D reconstruction methods using data gathered from a saline phantom. We found that without applying corrections to reconstructed images it was possible to estimate blood volume in a two-dimensional hemiarray case with an uncertainty of around 27 ml. In an approximately 3D hemiarray case, volume prediction was possible with a maximum uncertainty of around 38 ml in the centre of the electrode plane. After application of a QI normalizing filter, average uncertainties in a two-dimensional hemiarray case were reduced to about 15 ml. Uncertainties in the approximate 3D case were reduced to about 30 ml
Availability note (English)
Available from http://dx.doi.org/10.1088/0967-3334/29/8/005Additional details
Identifiers
- DOI
- 10.1088/0967-3334/29/8/005;
- PII
- S0967-3334(08)73460-9;
Publishing Information
- Journal Title
- Physiological Measurement (Print)
- Journal Volume
- 29
- Journal Issue
- 8
- Journal Page Range
- p. 913-927
- ISSN
- 0967-3334
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44127420
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- APPROXIMATIONS; BLOOD; COMPARATIVE EVALUATIONS; DECOMPOSITION; ELECTRODES; FILTERS; HEALTH HAZARDS; IMAGES; INJURIES; PATIENTS; PHANTOMS; SENSITIVITY; SPINAL CORD; TOMOGRAPHY; TWO-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- BIOLOGICAL MATERIALS; BODY FLUIDS; CALCULATION METHODS; CENTRAL NERVOUS SYSTEM; CHEMICAL REACTIONS; DIAGNOSTIC TECHNIQUES; DISEASES; EVALUATION; HAZARDS; MATERIALS; MOCKUP; NERVOUS SYSTEM; STRUCTURAL MODELS