Published June 2012
| Version v1
Journal article
Tissue compatibility inspection of central vein catheter sterilized by radiation
Creators
- 1. Wuzhong District CDC, Suzhou (China)
- 2. School of Radiation Medicine and Public Health, Medical College of Soochow University, Jiangsu Provincial Key Laboratory of Radiation Medicine and Protection, Suzhou (China)
Description
In order to inspect the security application of radiation sterilization medical products, the cytotoxicity, guinea pigs delayed contact sensitization and skin stimulation tests were carried out to evaluate the tissue compatibility of central vein catheter sterilized by irradiation and the radiation sterilization doses were set according to the ISO11137-2 criterion. The results show that the radiation sterilized samples had no cytotoxicity, no sensitization and no intracutaneous activation that meet the request of tissue compatibility for medical products. It suggests that the central vein catheter sterilized by radiation with set absorbed dose has good tissue compatibility. (authors)
Additional details
Publishing Information
- Journal Title
- Journal of Radiation Research and Radiation Processing
- Journal Volume
- 30
- Journal Issue
- 3
- Journal Page Range
- p. 165-169
- ISSN
- 1000-3436
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 51053352
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- ABSORBED RADIATION DOSES; ANIMAL TISSUES; COMPATIBILITY; GUINEA PIGS; RADIOSTERILIZATION; SKIN; TOXICITY; VEINS
- Descriptors DEC
- ANIMALS; BLOOD VESSELS; BODY; CARDIOVASCULAR SYSTEM; DOSES; IRRADIATION; MAMMALS; ORGANS; RADIATION DOSES; RODENTS; STERILIZATION; VERTEBRATES
Optional Information
- Notes
- 1 fig., 3 tabs., 10 refs.