Analysis of 3 dimension image
Description
An analysis was made of the geometric relation between stereo-radiographing system and its observation one which has an effect on the imaging in the diagnostic stereography. (1) In stereography with magnified X-rays in which the magnification ratio of images on X and Y axis is different from the one on Z axis a kind of image distortion is inevitable. (2) Stereographing system under a certain appropriate condition can produce an image with equal magnification ratio of one on X, Y and also Z giving life-size without distortion. (3) As described above, a magnified stereographic image accompanies a distortion. However, a magnified stereographic image will be perceived if the image on the retina is enlarged, which can be obtained by shortening the visual distance thus widening the visual angle. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 37
- Journal Issue
- 6
- Series
- Nippon Hoshasen Gijutsu Gakkai Zasshi.
- Journal Page Range
- 766-772
- ISSN
- 0369-4305
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 13703669
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; COMPUTERIZED SIMULATION; DISTANCE; EYES; FOCUSING; GEOMETRY; IMAGES; PHOTOGRAPHIC FILMS; X-RAY TUBES
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; ELECTRON TUBES; MATHEMATICS; MEDICINE; ORGANS; SENSE ORGANS; SIMULATION; X-RAY EQUIPMENT