Published November 1981 | Version v1
Journal article

Analysis of 3 dimension image

  • 1. Fukushima Medical Coll. (Japan)

Description

An analysis was made of the geometric relation between stereo-radiographing system and its observation one which has an effect on the imaging in the diagnostic stereography. (1) In stereography with magnified X-rays in which the magnification ratio of images on X and Y axis is different from the one on Z axis a kind of image distortion is inevitable. (2) Stereographing system under a certain appropriate condition can produce an image with equal magnification ratio of one on X, Y and also Z giving life-size without distortion. (3) As described above, a magnified stereographic image accompanies a distortion. However, a magnified stereographic image will be perceived if the image on the retina is enlarged, which can be obtained by shortening the visual distance thus widening the visual angle. (author)

Additional details

Publishing Information

Journal Title
Nippon Hoshasen Gijutsu Gakkai Zasshi
Journal Volume
37
Journal Issue
6
Series
Nippon Hoshasen Gijutsu Gakkai Zasshi.
Journal Page Range
766-772