Published 1987
| Version v1
Journal article
Dependence on the gamma radiation energy of the NaI(Tl) scintillation detector efficiency
- 1. Institutul de Fizica si Inginerie Nucleara, Bucharest (Romania)
- 2. Combinatul Siderurgic Galati. Galati, Romania
Description
A mathematical method, based on Monte Carlo procedure, has been elaborated in order to determine the dependence of the NaI(Tl) scintillation detector efficiency on the gamma radiation energy. The calculus program elaborated for an isotropic point source takes into account the other parameters the detector efficiency depends on: the NaI(Tl) scintillation crystal dimensions and the source-detector distance. The result obtained by Monte Carlo Method, for two NaI(Tl) crystals of different dimensions are compared with the results given in the literature and with those ones obtained experimentally. A good agreement of the results can be observed.(authors)
Additional details
Additional titles
- Original title (Romanian)
- Dependenta eficacitatii detectorurului de scintilatie cu scintilator NaI(Tl) de energia radiatiei gamma
Publishing Information
- Journal Title
- Studii si Cercetari de Fizica
- Journal Volume
- 39
- Journal Issue
- 4
- Series
- Stud. Cercet. Fiz.
- Journal Page Range
- 353-356
- ISSN
- 0039-3940
- CODEN
- SCEFA
INIS
- Country of Publication
- Romania
- Country of Input or Organization
- Romania
- INIS RN
- 20027081
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- GAMMA DETECTION; MONTE CARLO METHOD; SCINTILLATION COUNTERS; SENSITIVITY
- Descriptors DEC
- DETECTION; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS