Published October 2006 | Version v1
Journal article

Data acquisition ATCA system for neutron and gamma-rays spectrometries

  • 1. Associacao EURATOM/IST Centro de Fusao Nuclear, Instituto Superior Tecnico, 1049-001 Lisbon (Portugal)

Description

Digital pulse processing (DPP) systems are known to have better performance than analog ones for neutron and/or gamma-ray pulses. DPP can synthesize almost any pulse response shape without the associated signal degradation which happens in a complex analog path. Measuring techniques involving detectors/spectrometers for fusion diagnostics rely on real-time algorithms, implemented in field programmable gate array (FPGA), for pulse height analysis, pulse shape discrimination, and pileup rejection of digitized pulses in real time for reduced data throughput, monitoring, and control. This article describes a data acquisition system for real-time pulse analysis based on ATCA and contains a 6 GFLPOS ix86-based control unit and a number of transient recorder (TR) modules interconnected through PCI Express links. Each TR module features (i) eight channels of 12 bit resolution with accuracy equal or higher than 10 bits, (ii) 200 Msamples/s of sampling rate achieving 400 Msamples/s in an interleaved architecture, (iii) 2 or 4 Gbytes of local memory, and (iv) two field FPGAs able to perform real-time processing algorithms

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
77
Journal Issue
10
Journal Page Range
p. 10F511
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2006 American Institute of Physics