Published June 1, 1991 | Version v1
Journal article

Calculation of the time dependence of the pulse shape in a cylindrical proportional counter using the Diethorn assumptions

  • 1. Giessen Univ. (Germany, F.R.). 1. Physikalisches Inst.

Description

A recently developed equipment for the readout of TSEE dosemeters employs a wide-range proportional counter for exoelectron detection. In order to improve its pulse resolution we were interested in the exact contribution of electrons to the time dependence of the pulse. This contribution is completely neglected by the Ramsey-Wilkinson theory. Therefore the pulse shape was calculated by using the assumption of the Diethorn approximation, taking the electron contribution into account. With respect to the ion contribution the results showed good agreement with the Ramsey-Wilkinson theory. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
303
Journal Issue
2
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
370-373
ISSN
0168-9002
CODEN
NIMAE