Published 2004 | Version v1
Miscellaneous

Selection of anode-cathode distance in x-ray tube manufacturing

Description

Factors affecting the choice of anode-cathode distance in x-ray tube are reviewed. Among these factors the most important one is found to be the sparking voltage. Through solving the paraxial ray equation and accepting 0.1 mm tolerance in focal spot size, a relationship between maximum anode voltage and Anode-Cathode distance is obtained

Availability note (English)

Available from Atomic Energy Organization of Iran

Additional details

Publishing Information

Publisher
Amirkabir University of Technology
Imprint Place
Tehran, On (Iran, Islamic Republic of)
Imprint Pagination
[3 p.]

Conference

Title
International Conference on Physics
Dates
6-9 Jan 2004
Place
Tehran (Iran, Islamic Republic of)

INIS

Country of Publication
Iran, Islamic Republic of
Country of Input or Organization
Iran, Islamic Republic of
INIS RN
35104923
Subject category
S42: ENGINEERING;
Resource subtype / Literary indicator
Conference, Non-conventional Literature
Descriptors DEI
ANODES; CATHODES; DESIGN; ELECTRIC POTENTIAL; ELECTRIC SPARKS; MANUFACTURING; SPARK GAPS; SPECIFICATIONS; X-RAY TUBES
Descriptors DEC
ELECTRIC DISCHARGES; ELECTRODES; ELECTRON TUBES; EQUIPMENT; X-RAY EQUIPMENT