Published March 2007
| Version v1
Journal article
Feedback system for improving the performance of extended x-ray absorption fine structure measurements
Creators
- 1. SPring-8/Japan Synchrotron Radiation Research Institute, Sayo-cho, Sayo-gun, Hyogo 679-5198 (Japan)
Description
A rapid fine energy tuning method for a double-crystal monochromator (DCM) is developed by combining a feedback control system with a phase sensitive detector. This method maintains both a parallel angle between the first and second crystals of the DCM and the peak of the rocking curve. The method is used to perform a continuous energy scan of the DCM without readjustment of the angle between the two crystals. The time taken by this method to measure the extended x-ray absorption fine structure is 60% of that taken by a conventional procedure
Additional details
Identifiers
- DOI
- 10.1063/1.2713436;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 78
- Journal Issue
- 3
- Journal Page Range
- p. 033104-033104.4
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38104663
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Descriptors DEI
- ABSORPTION; ABSORPTION SPECTROSCOPY; CONTROL SYSTEMS; CRYSTALS; FEEDBACK; FINE STRUCTURE; NEUTRON DIFFRACTION; PEAKS; PERFORMANCE; X RADIATION; X-RAY SPECTROSCOPY
- Descriptors DEC
- COHERENT SCATTERING; DIFFRACTION; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; RADIATIONS; SCATTERING; SORPTION; SPECTROSCOPY
Optional Information
- Notes
- (c) 2007 American Institute of Physics