Establishment of a procedure for calibrating neutron monitors at the Physics Institute of the University of Sao Paulo, Brazil
Description
A facility for neutron irradiation was designed and built in order to apply the ISO draft procedures for calibrating and determining the response of neutron measuring devices to Brazilian neutron monitors and dosemeters used for radiation protection. As a test for the system, the three methods proposed by the draft were applied. A calibrated Am-Be source, a Li(Eu) sensor, a Bonner multisphere detector, and three shadow cones were used. Based on the data from the Bonner spheres, the free field fluence response, the fractional room scattering contribution at unit calibration distance, and the source detector characteristic constant were determined. The unfolded neutron spectrum was also determined and the corresponding fluence, dose equivalent, and mean energy were calculated. All of them proved to be in good agreement with values expected from the spectrum of the neutron source. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 44
- Journal Issue
- 1-4
- Journal Page Range
- p. 143-146.
- ISSN
- 0144-8420
- CODEN
- RPDODE
Conference
- Title
- 7. Symposium on neutron dosimetry.
- Dates
- 14-18 Oct 1991.
- Place
- Berlin (Germany).
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 24038344
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BONNER SPHERE DETECTORS; BRAZIL; CALIBRATION; NEUTRON MONITORS; NEUTRON SPECTRA
- Descriptors DEC
- DEVELOPING COUNTRIES; LATIN AMERICA; MEASURING INSTRUMENTS; MODERATING DETECTORS; MONITORS; NEUTRON DETECTORS; RADIATION DETECTORS; RADIATION MONITORS; SOUTH AMERICA; SPECTRA