10B coated multigrid position sensitive neutron detector with zero parallax
- 1. Solid State Physics Division, Bhabha Atomic Research Centre, Mumbai (India)
Description
10B as a solid state converter material for neutron detection is being revisited due to its properties such as chemically non reactive, nontoxic and high energy reaction products. The disadvantage is low efficiency as only thin layer can be used for the detection, where coating thickness should be less than range of 0.91 MeV Li. Gain in efficiency can be attained using two strategies: multiple stacking of thin layers or by inclining the converter layer to increase effective neutron penetration depth as in Jalousie detector. Consideration on design of position sensitive detector (PSD) is for the 10B layer thickness, detection efficiency and transmission of neutron beam to following consecutive 10B layers. Design of PSD with stacking of multiple 10B layers, for improvement in efficiency, involves complex anode-cathode geometry. It results in broadening of peak due to parallax. This paper deals with the novel design and fabrication of a PSD based on multilayer 10B, multiple anode grids and zero parallax. It is essential to handle skilfully hardware design, in addition to complexity of data acquisition electronics. The structure follows the curvature equivalent to sample to detector distance as radius. A PSD consists of four 10B coated plates and an array of three multiwire grids arranged alternately in curved casing. Incident neutron beam thus intercepts six 10B coated surfaces. An assembled curvilinear PSD and the layout of teflon holder supporting curved anode-cathode structure is shown. Position readout is carried out using charge division encoding method. An arc length of 0.7 m (detector sensitive length) covers the angular range of 23°. Central anode grid is mounted at radius 2000 mm and other two grids at ± 12 mm radii from central grid position. At constant sensitive length of 700 mm for all three grids, deviation in 2 for other two grids is ± 0.12 ∼ ± 4 mm, when central anode wires on multiwire grids are aligned. This deviation in angular position can be normalized with linear position using a software correction. Presently PSD with single grid is tested and position resolution obtained is 5 mm∼ anode pitch. Overview of these developments with multiple grids is presented.
Additional details
Publishing Information
- Publisher
- Bhabha Atomic Research Centre
- Imprint Place
- Mumbai (India)
- Imprint Title
- Proceedings of the sixth conference on neutron scattering: programme and abstract
- Imprint Pagination
- 178 p.
- Journal Page Range
- p. 86
Conference
- Title
- 6. conference on neutron scattering
- Acronym
- CNS2016
- Dates
- 21-23 Nov 2016
- Place
- Mumbai (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 48017267
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BORON 10 BEAMS; DATA ACQUISITION SYSTEMS; DESIGN; NEUTRON DETECTORS; POSITION SENSITIVE DETECTORS
- Descriptors DEC
- BEAMS; ION BEAMS; MEASURING INSTRUMENTS; RADIATION DETECTORS
Optional Information
- Notes
- 3 refs., 1 fig.