Description

The feasibility of heavy ion computed tomography was investigated. A cylindrical polyethylene sample of 50 mm in diameter was irradiated with 290 MeV/u and 400 MeV/u carbon ions and 150 MeV/u helium ions. The CT images were reconstructed by moving the sample only in the transverse direction because of its symmetrical structure. A Lucite sample with many holes of different diameters was also irradiated. From these images the spatial resolution of this CT system was estimated to be less than 2 mm. The electron density ratios of ethyl alcohol, water, and Lucite to polyethylene were obtained, which are in good agreement with the calculated values. (author)

Part of:
Proceedings of the second symposium on accelerator and related technology for application

Additional details

Publishing Information

Imprint Title
Proceedings of the second symposium on accelerator and related technology for application
Imprint Pagination
129 p.
Journal Page Range
p. 17-20

Conference

Title
2. symposium on accelerator and related technology for application
Acronym
ARTA 1999
Dates
1-3 Dec 1999
Place
Tokyo (Japan)

Optional Information