Tissue-equivalent torso phantom for calibration of transuranic-nuclide counting facilities

Description

Several tissue-equivalent human-torso phantoms have been constructed for the calibration of counting systems used for in-vivo measurement of transuranic radionuclides. The phantoms contain a simulated human rib cage (in some cases, real bone) and removable model organs, and they include tissue-equivalent chest plates that can be placed over the torso to simulate people with a wide range of statures. The organs included are the lungs, liver, and tracheobronchial lymph nodes. Polyurethane with varying concentrations of added calcium was used to simulate the linear photon-attenuation properties of various human tissues, including lean muscle, adipose-muscle mixtures, cartilage, and bone. Foamed polyurethane was used to simulate lung tissue. Organs have been loaded with highly pure 238Pu, 239Pu, 241Am, and other radionuclides of interest. The validity of the phantom as a calibration standard has been checked in separate intercomparison studies using human subjects whose lungs contained a plutonium simulant. The resulting phantom calibration factors generally compared to within +-20% of the average calibration factors obtained for the human subjects

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01; 1 as DE86009693.

Files

17078054.pdf

Files (309.3 kB)

Name Size Download all
md5:098799e9db20b342ddbd39f13f9ada51
309.3 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
23 p.
Report number
UCRL--93776

Conference

Title
Radiobioassay and internal dosimetry workshop.
Dates
20 Jan 1986.
Place
Albuquerque, NM (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17078054
Subject category
S07: ISOTOPES AND RADIATION SOURCES;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; CHEST; IN VIVO; MAN; PHANTOMS; RADIATION DETECTORS; TISSUE-EQUIVALENT MATERIALS; TRANSURANIUM ELEMENTS
Descriptors DEC
ANIMALS; BODY; BODY AREAS; ELEMENTS; MAMMALS; MATERIALS; MEASURING INSTRUMENTS; MOCKUP; PRIMATES; STRUCTURAL MODELS; VERTEBRATES

Optional Information

Notes
Portions of this document are illegible in microfiche products.
Secondary number(s)
CONF-860146--8.