Tissue-equivalent torso phantom for calibration of transuranic-nuclide counting facilities
Description
Several tissue-equivalent human-torso phantoms have been constructed for the calibration of counting systems used for in-vivo measurement of transuranic radionuclides. The phantoms contain a simulated human rib cage (in some cases, real bone) and removable model organs, and they include tissue-equivalent chest plates that can be placed over the torso to simulate people with a wide range of statures. The organs included are the lungs, liver, and tracheobronchial lymph nodes. Polyurethane with varying concentrations of added calcium was used to simulate the linear photon-attenuation properties of various human tissues, including lean muscle, adipose-muscle mixtures, cartilage, and bone. Foamed polyurethane was used to simulate lung tissue. Organs have been loaded with highly pure 238Pu, 239Pu, 241Am, and other radionuclides of interest. The validity of the phantom as a calibration standard has been checked in separate intercomparison studies using human subjects whose lungs contained a plutonium simulant. The resulting phantom calibration factors generally compared to within +-20% of the average calibration factors obtained for the human subjects
Availability note (English)
MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01; 1 as DE86009693.Files
17078054.pdf
Files
(309.3 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:098799e9db20b342ddbd39f13f9ada51
|
309.3 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 23 p.
- Report number
- UCRL--93776
Conference
- Title
- Radiobioassay and internal dosimetry workshop.
- Dates
- 20 Jan 1986.
- Place
- Albuquerque, NM (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17078054
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; CHEST; IN VIVO; MAN; PHANTOMS; RADIATION DETECTORS; TISSUE-EQUIVALENT MATERIALS; TRANSURANIUM ELEMENTS
- Descriptors DEC
- ANIMALS; BODY; BODY AREAS; ELEMENTS; MAMMALS; MATERIALS; MEASURING INSTRUMENTS; MOCKUP; PRIMATES; STRUCTURAL MODELS; VERTEBRATES
Optional Information
- Notes
- Portions of this document are illegible in microfiche products.
- Secondary number(s)
- CONF-860146--8.