Use of active personal dosemeters in interventional radiology and cardiology: Tests in hospitals – ORAMED project
Creators
- 1. Belgian Nuclear Research Centre (SCK-CEN), Radiation Protection Dosimetry and Calibration, Boeretang 200, 2400 Mol (Belgium)
- 2. Greek Atomic Energy Commission (GAEC), Ag. Paraskevi, 15310 Athens (Greece)
- 3. Institut de Radioprotection et de Sûreté Nucléaire (IRSN), BP17, 92262 Fontenay-aux-Roses (France)
- 4. Universitat Politècnica de Catalunya (UPC), Diagonal 647, 08028 Barcelona (Spain)
- 5. Hogeschool-Universiteit Brussel (HUB), Blekerijstraat 23-29, 1000 Brussel (Belgium)
- 6. Institut Universitaire de Radiophysique Appliquée (IRA), Grand Pré 1, 1007 Lausanne (Switzerland)
Description
Although active personal dosemeters (APDs) are not used quite often in hospital environments, the possibility to assess the dose and/or dose rate in real time is particularly interesting in interventional radiology and cardiology (IR/IC) since operators can receive relatively high doses while standing close to the primary radiation field. A study concerning the optimization of the use of APDs in IR/IC was performed in the framework of the ORAMED project, a Collaborative Project (2008–2011) supported by the European Commission within its 7th Framework Program. This paper reports on tests performed with APDs on phantoms using an X-ray facility in a hospital environment and APDs worn by interventionalists during routine practice in different European hospitals. The behaviour of the APDs is more satisfactory in hospitals than in laboratories with respect to the influence of the tube peak high voltage and pulse width, because the APDs are tested in scattered fields with dose equivalent rates generally lower than 1 Sv.h−1. .
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radmeas.2011.08.002Additional details
Identifiers
- DOI
- 10.1016/j.radmeas.2011.08.002;
- PII
- S1350-4487(11)00385-4;
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 46
- Journal Issue
- 11
- Journal Page Range
- p. 1258-1261
- ISSN
- 1350-4487
- CODEN
- RMEAEP
Conference
- Title
- International workshop on optimization of radiation protection of medical staff
- Acronym
- ORAMED 2011
- Dates
- 20-22 Jan 2011
- Place
- Barcelona (Spain)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 44087411
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DOSE EQUIVALENTS; DOSE RATES; DOSEMETERS; HOSPITALS; MEDICAL PERSONNEL; OPTIMIZATION; PERSONNEL DOSIMETRY; PERSONNEL MONITORING; PHANTOMS; RADIOLOGY
- Descriptors DEC
- BUILDINGS; DOSIMETRY; MEASURING INSTRUMENTS; MEDICAL ESTABLISHMENTS; MEDICINE; MOCKUP; MONITORING; NUCLEAR MEDICINE; PERSONNEL; RADIATION MONITORING; STRUCTURAL MODELS