A low-cost liquid refractive index sensor based on plastic optical fibre and CCD array

  • 1. Instituto de Telecomunicações—Polo de Aveiro, Aveiro 3810-193 (Portugal)

Description

A wireless refractive index sensor for in-line measurement in liquids was developed. It is based on the Fresnel reflection on a prism–liquid interface, where the critical angle of the totally reflected light in the interface of two media is detected by a charge-coupled device (CCD) array. A LED is used as a light source and a plastic optical fibre, with a well-established numeric aperture, guides the light to the prism interface. Both the LED and the CCD are controlled with a Raspberry Pi 3 Model B board with wireless communication. Resolutions with values lower than 1.0  ×  10−4 RIU can be achieved with this potential low-cost configuration. (technical note)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6501/ab46ff

Additional details

Identifiers

Publishing Information

Journal Title
Measurement Science and Technology
Journal Volume
31
Journal Issue
4
Journal Page Range
[7 p.]
ISSN
0957-0233
CODEN
MSTCEP

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
52117628
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CHARGE-COUPLED DEVICES; INTERFACES; LIQUIDS; OPTICAL FIBERS; REFRACTIVE INDEX; SENSORS
Descriptors DEC
FIBERS; FLUIDS; OPTICAL PROPERTIES; PHYSICAL PROPERTIES; SEMICONDUCTOR DEVICES