Dynamic flat panel detector versus image intensifier in cardiac imaging: dose and image quality

  • 1. Radiology Department, Medicine School, Complutense University and San Carlos University Hospital, 28040 Madrid (Spain)
  • 2. Siemens AG, Medical Solutions, D-91301 Forchheim (Germany)
  • 3. Radiation Protection Division, Ministry of Health, L-2120 (Luxembourg)
  • 4. National Institute of Cardiac Surgery and Interventional Cardiology (INCCI), L-2120 (Luxembourg)

Description

The practical aspects of the dosimetric and imaging performance of a digital x-ray system for cardiology procedures were evaluated. The system was configured with an image intensifier (II) and later upgraded to a dynamic flat panel detector (FD). Entrance surface air kerma (ESAK) to phantoms of 16, 20, 24 and 28 cm of polymethyl methacrylate (PMMA) and the image quality of a test object were measured. Images were evaluated directly on the monitor and with numerical methods (noise and signal-to-noise ratio). Information contained in the DICOM header for dosimetry audit purposes was also tested. ESAK values per frame (or kerma rate) for the most commonly used cine and fluoroscopy modes for different PMMA thicknesses and for field sizes of 17 and 23 cm for II, and 20 and 25 cm for FD, produced similar results in the evaluated system with both technologies, ranging between 19 and 589 μGy/frame (cine) and 5 and 95 mGy min-1 (fluoroscopy). Image quality for these dose settings was better for the FD version. The 'study dosimetric report' is comprehensive, and its numerical content is sufficiently accurate. There is potential in the future to set those systems with dynamic FD to lower doses than are possible in the current II versions, especially for digital cine runs, or to benefit from improved image quality

Availability note (English)

Available online at http://stacks.iop.org/0031-9155/50/5731/pmb5_23_022.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
50
Journal Issue
23
Journal Page Range
p. 5731-5742
ISSN
0031-9155
CODEN
PHMBA7